STATIC AND DYNAMIC COMPLIANCE 22 777404-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2016-05-24 for STATIC AND DYNAMIC COMPLIANCE 22 777404-101 manufactured by Carefusion, Inc.

Event Text Entries

[45815147] (b)(4). Results of investigation: the carefusion failure analysis lab received the device and performed a failure investigation. The customer's reported issue was duplicated and it was also noted that the power supply had a burning component once turned on.
Patient Sequence No: 1, Text Type: N, H10


[45815148] The customer states that she is getting leak errors and there is a squealing noise when inhaling oxygen for nitrogen washout test. It is unknown whether there was any patient involvement. It was noted on the failure analysis that the power supply had a burning component once turned on.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2031702-2016-00650
MDR Report Key5676389
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2016-05-24
Date of Report2015-02-18
Date of Event2015-02-18
Date Mfgr Received2016-05-02
Date Added to Maude2016-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149227830
Manufacturer G1CAREFUSION, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2016-05-24
Returned To Mfg2015-03-13
Model Number22
Catalog Number777404-101
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-24