MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-25 for RELIANCE ENDOSCOPE PROCESSING SYSTEM manufactured by Steris Canada Corporation.
[45976188]
The user facility also reported that their processor experienced a fault 36: filter pressure too low. A steris service technician inspected the processor and found that a water leak from the connection block and thermocouple compression had dripped down onto the air compressor. The water making contact with the air compressor caused the component to not operate properly and emit the burning smell as reported by the user facility. When water made contact with the air compressor, the compressor could not generate enough pressure thus causing the fault 36. The technician repaired the processor, ran a test cycle, and confirmed the unit to be operating properly. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10
[45976189]
The user facility reported that a burning smell was emitting from their processor. No report of injury or procedural delays or cancellations.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 9680353-2016-00051 |
MDR Report Key | 5679869 |
Date Received | 2016-05-25 |
Date of Report | 2016-05-25 |
Date of Event | 2016-04-27 |
Date Mfgr Received | 2016-04-27 |
Date Added to Maude | 2016-05-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. KATHRYN CADORETTE |
Manufacturer Street | 5960 HEISLEY ROAD |
Manufacturer City | MENTOR OH 44060 |
Manufacturer Country | US |
Manufacturer Postal | 44060 |
Manufacturer Phone | 4403927231 |
Manufacturer G1 | STERIS CANADA CORPORATION |
Manufacturer Street | 490, ARMAND-PARIS |
Manufacturer City | QUEBEC, QUEBEC GIC 8A3 |
Manufacturer Country | CA |
Manufacturer Postal Code | GIC 8A3 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | RELIANCE ENDOSCOPE PROCESSING SYSTEM |
Generic Name | PROCESSING SYSTEM |
Product Code | NZA |
Date Received | 2016-05-25 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STERIS CANADA CORPORATION |
Manufacturer Address | 490 ARMAND-PARIS QUEBEC, QUEBEC GIC 8A3 CA GIC 8A3 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2016-05-25 |