Event Text Entries

[45957660] There is no defect or malfunction of the device associated with this event. All pertinent information available to second sight medical products has been submitted. The company is submitting this mdr to ensure full compliance with 21 cfr part 803.
Patient Sequence No: 1, Text Type: N, H10


[45957661] Patient (b)(6) was implanted with the argus ii device on (b)(6) 2016. One week post-op, it was reported that this patient experienced high intraocular pressure (44 mmhg) in the implanted eye. The patient was administered anti-glaucoma medication. The root cause for the high intraocular pressure was determined to be a tight scleral band. The patient underwent a surgical intervention on (b)(6) 2016 during which the surgeon loosened the scleral band. On (b)(6) 2016, the surgeon reported that the patient's intraocular pressure had dropped to 24 mmhg.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004081696-2016-00007
MDR Report Key5680641
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2016-05-25
Date of Report2016-04-27
Date of Event2016-04-26
Date Mfgr Received2016-04-27
Device Manufacturer Date2014-06-02
Date Added to Maude2016-05-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MARK LOAR
Manufacturer Street12744 SAN FERNANDO RD. SUITE 400
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal91342
Manufacturer Phone8188335049
Manufacturer G1SECOND SIGHT MEDICAL PRODUCTS, INC.
Manufacturer Street12744 SAN FERNANDO RD. SUITE 400
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal Code91342
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARGUS II RETINAL PROSTHESIS
Generic NameRETINAL PROSTHESIS
Product CodeNBF
Date Received2016-05-25
Model Number011014-001-K
Catalog Number011014
Lot NumberN/A
Device Expiration Date2016-06-02
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSECOND SIGHT MEDICAL PRODUCTS, INC.
Manufacturer Address12744 SAN FERNANDO RD. SUITE 400 SYLMAR CA 91342 US 91342


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2016-05-25