STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 29C 777409-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-05-26 for STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 29C 777409-101 manufactured by Carefusion, Inc.

Event Text Entries

[46010327] (b)(4) results of investigation: the device was evaluated by a carefusion field service engineer on-site. He was able to duplicate the problem and it was resolved by upgrading the firmware from version 2. 1 to 2. 2 and adjusting the timeout settings. The device is currently working to service specifications with no parts replaced or returned.
Patient Sequence No: 1, Text Type: N, H10


[46010328] The customer reported her system has intermittent graphics issues as well as the treadmill shut off in middle of test. She was running an exercise test on a female, young adult. There was no harm to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2016-03797
MDR Report Key5682231
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-05-26
Date of Report2016-05-04
Date of Event2016-05-04
Date Mfgr Received2016-05-04
Device Manufacturer Date2012-10-01
Date Added to Maude2016-05-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149227830
Manufacturer G1CAREFUSION, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2016-05-26
Model NumberVMAX ENCORE 29C
Catalog Number777409-101
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-05-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.