INTELLIGENT INFUSOR 2001 UNK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1996-12-13 for INTELLIGENT INFUSOR 2001 UNK manufactured by Mcgaw Medical.

Event Text Entries

[36019] Ob pt was started on iv pitocin controlled by iv controller set at 3ml/hr by pt's rn. Ran 16 mins, controller then alarmed "repair 3". Second rn responded by setting controller to "hold" and attended to pt. First rn arrived and noted pitocin infusing at "wide open condition. " unknown amount of medication delivered. Rn clamped line, treated pt with o2, position change and iv flush with lactated ringers solution for uterine tetany and fhr deceleration. Delivered 24 hrs later. No known adverse affect on mother or baby.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number56856
MDR Report Key56856
Date Received1996-12-13
Date of Report1996-05-24
Date of Event1996-05-13
Date Facility Aware1996-05-14
Report Date1996-05-24
Date Reported to Mfgr1996-05-24
Date Added to Maude1996-12-20
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameINTELLIGENT INFUSOR
Generic NameINFUSION CONTROLLER
Product CodeLDR
Date Received1996-12-13
Model Number2001
Catalog NumberUNK
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeUNKNOWN
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key57373
ManufacturerMCGAW MEDICAL
Manufacturer Address1601 WALLACE DRIVE CARROLLTON TX 75006 US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 1996-12-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.