MAUDE MDR 5689631

MDR report key
5689631
Report number
1419185-2016-00002
Event key
0
Event type
3
Date of event
2015-04-23
Date received
2016-05-31
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
MS JESSICA BARRILE
Address
1111 LAKESIDE DR GURNEE IL 60031 US
Phone
847-847-8478
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1PUSH TO SET INTERMITTENT SUCTION REGULATORPTS ISUOHIO MEDICALKDP12518701-1251-908ISU077016Y R

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12016-05-310

Event Narratives#

N

Patient 1

D

Patient 1

A USER FACILITY REPORTED TO US THAT THE DEVICE "WONT CYCLE OFF" ON (B)(6)2015. THERE WAS NO PATIENT INVOLVEMENT, AND THE INCIDENT DID NOT RESULT IN A DELAY OF TREATMENT. THE DEVICE WAS EVALUATED ON (B)(6) 2015 AND IT WAS NOTED THAT THE DEVICE HAD A DAMAGED DIAPHRAGM. OUR RISK DOCUMENT CLASSIFIES THE SEVERITY OF A DAMAGED DIAPHRAGM AS "POTENTIALLY REQUIRING MEDICAL INTERVENTION TO PREVENT IMPAIRMENT OF A BODY FUNCTION OR PERMANENT DAMAGE TO A BODY STRUCTURE". WE ARE OBTAINING FURTHER CLINICAL EVALUATION OF THE RISK OF THIS FAILURE AND WILL PROVIDE FOLLOW UP TO THIS REPORT ONCE OBTAINED.

N

Patient 1

THIS IS A CORRECTION TO REPORT SUBMITTED ON 01/29/2016: A USER FACILITY REPORTED TO US THAT THE DEVICE "GOES FROM FULL TO LOW, HARD TO ADJUST" ON (B)(6) 2015. THERE WAS NO PATIENT INVOLVEMENT, AND THE INCIDENT DID NOT RESULT IN DELAY OF TREATMENT. THE DEVICE WAS EVALUATED ON 04/23/2015 AND IT WAS NOTED THAT THE DEVICE HAD A DAMAGED DIAPHRAGM. OUR RISK DOCUMENT CLASSIFIES THE SEVERITY OF A DAMAGED DIAPHRAGM AS "POTENTIALLY REQUIRING MEDICAL INTERVENTION TO PREVENT PERMANENT IMPAIRMENT OF A BODY FUNCTION OR PERMANENT IMPAIRMENT OF A BODY FUNCTION OR PERMANENT DAMAGE TO A BODY STRUCTURE". WE ARE OBTAINING FURTHER CLINICAL EVALUATION OF THE RISK OF THIS FAILURE AND WILL PROVIDE FOLLOW UP TO THIS REPORT ONCE OBTAINED.

D

Patient 1

THIS IS A CORRECTION TO REPORT SUBMITTED ON 01/29/2016: A USER FACILITY REPORTED TO US THAT THE DEVICE "GOES FROM FULL TO LOW, HARD TO ADJUST" ON (B)(6) 2015. THERE WAS NO PATIENT INVOLVEMENT, AND THE INCIDENT DID NOT RESULT IN DELAY OF TREATMENT. THE DEVICE WAS EVALUATED ON 04/23/2015 AND IT WAS NOTED THAT THE DEVICE HAD A DAMAGED DIAPHRAGM. OUR RISK DOCUMENT CLASSIFIES THE SEVERITY OF A DAMAGED DIAPHRAGM AS "POTENTIALLY REQUIRING MEDICAL INTERVENTION TO PREVENT PERMANENT IMPAIRMENT OF A BODY FUNCTION OR PERMANENT IMPAIRMENT OF A BODY FUNCTION OR PERMANENT DAMAGE TO A BODY STRUCTURE". WE ARE OBTAINING FURTHER CLINICAL EVALUATION OF THE RISK OF THIS FAILURE AND WILL PROVIDE FOLLOW UP TO THIS REPORT ONCE OBTAINED.

N

Patient 1

CLINICAL ASSESSMENT OF THE FAILURE MODE "DIFFICULT TO ADJUST" HAS BEEN OBTAINED AND IT HAS BEEN DETERMINED THAT THIS OCCURRENCE DOES NOT REPRESENT A REPORTABLE MALFUNCTION DUE TO NO EVIDENCE EXISTING TO REASONABLY SUGGEST THAT THE MALFUNCTION WOULD BE LIKELY TO CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY IF IT WERE TO RECUR. (B)(4).

D

Patient 1

CLINICAL ASSESSMENT OF THE FAILURE MODE "DIFFICULT TO ADJUST" HAS BEEN OBTAINED AND IT HAS BEEN DETERMINED THAT THIS OCCURRENCE DOES NOT REPRESENT A REPORTABLE MALFUNCTION DUE TO NO EVIDENCE EXISTING TO REASONABLY SUGGEST THAT THE MALFUNCTION WOULD BE LIKELY TO CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY IF IT WERE TO RECUR.