CARE-E-VAC 3 758000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-31 for CARE-E-VAC 3 758000 manufactured by Ohio Medical Llc.

Event Text Entries

[46918993]
Patient Sequence No: 1, Text Type: N, H10


[46918994] Ohio medical received customer complaint for portable suction device on march 31, 2016 stating "burn spot on board". Device was received by manufacturer on 4/13/2016, it was noted on 4/19/2016 that board (power supply) did in fact exhibit damage. Upon further evaluation, it was noted that device was contaminated by fluid ingress. It is suspected that fluid came into contact with power supply component, causing the burn/ damage. Ohio medical service manual states the following: page 5, processing & cleaning section 2. "do not allow any cleaning solution to spill into the vents on the unit. " and page 7 d) warning: decontamination 4. B. "items that must be replaced if exposed to fluids and contaminates: tubing, control circuit, power supply, pump, switch. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419185-2016-00026
MDR Report Key5689896
Date Received2016-05-31
Date of Report2016-04-29
Date of Event2016-03-31
Date Mfgr Received2016-03-31
Date Added to Maude2016-05-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS JESSICA BARRILE
Manufacturer Street1111 LAKESIDE DR
Manufacturer CityGURNEE IL 60031
Manufacturer CountryUS
Manufacturer Postal60031
Manufacturer Phone8478556318
Manufacturer G1OHIO MEDICAL LLC
Manufacturer Street1111 LAKESIDE DR
Manufacturer CityGURNEE IL 60031
Manufacturer CountryUS
Manufacturer Postal Code60031
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCARE-E-VAC 3
Generic NamePORTABLE SUCTION DEVICE
Product CodeBTA
Date Received2016-05-31
Returned To Mfg2016-04-13
Model Number758000
Catalog Number758000
Lot NumberCEB141017
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOHIO MEDICAL LLC
Manufacturer Address1111 LAKESIDE DR GURNEE IL 60031 US 60031


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-31