55" MEDICAL GRADE WALL MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-06-01 for 55" MEDICAL GRADE WALL MONITOR manufactured by Vts Medical Systems.

Event Text Entries

[47354983] The wall monitor subject of the event was located outside of the sterile field and is used to display duplicate images from other monitors such as patient data or case status. The wall monitor is not manufactured by steris and is manufactured by a third party. The monitor was returned to the oem for evaluation. The manufacturer evaluated the monitor and found that the power supply was not functioning. Further evaluation determined that there was an internal short between the positive and negative terminals of the ac input causing the elements to melt. It was also found that the circuit protection fuses were damaged. The short may have been caused by a power surge at the facility or a liquid leaking into the monitor. The manufacturer replaced the power supply and is returning the monitor to steris. The user facility received a replacement monitor and no further issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[47354984] The user facility reported that their wall monitor began to spark and smoke. No report of injury or procedural delays or cancellations.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000404456-2016-00005
MDR Report Key5692317
Date Received2016-06-01
Date of Report2016-06-01
Date of Event2016-05-04
Date Mfgr Received2016-05-04
Date Added to Maude2016-06-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1VTS MEDICAL SYSTEMS
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name55" MEDICAL GRADE WALL MONITOR
Generic NameWALL MONITOR
Product CodeKQM
Date Received2016-06-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVTS MEDICAL SYSTEMS
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.