Event Text Entries

[47649661] Product was not returned to the manufacturer for evaluation. Therefore, we are unable to determine the definitive cause of the reported event. However, applicable imaging studies were returned to the manufacturer for evaluation, whose review is as follows: "l5-s1 construct screws show probable bilateral s1 screw breakage. A spondylolisthesis is present. There is no interbody graft. It is unknown what the degree of reduction of the spondylolisthesis was immediately post-op. An interbody graft may help with deformity reduction and fusion. Fusion status is unknown. Root cause indeterminate. "
Patient Sequence No: 1, Text Type: N, H10


[47649662] It was reported that the patient underwent spinal fusion from l5 to s1 due to spondylolisthesis (grade 1) / spinal stenosis. Post-op, on an unknown date, (after 3 months), both the s1 level screws were visibly broken at the 4th or 5th thread. Patient was currently doing physiotherapy. The screws were still inside the patient. Reportedly, the patient was overall doing fine.
Patient Sequence No: 1, Text Type: D, B5


[100604595] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1030489-2016-01616
MDR Report Key5694472
Date Received2016-06-02
Date of Report2016-05-04
Date Mfgr Received2016-05-04
Date Added to Maude2016-06-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactGREG ANGLIN
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal38132
Manufacturer Phone9013963133
Manufacturer G1MEDTRONIC SOFAMOR DANEK
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal Code38132
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCD HORIZON SPINAL SYSTEM
Product CodeOSH
Date Received2016-06-02
Model NumberNA
Catalog NumberUNK
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC SOFAMOR DANEK
Manufacturer Address1800 PYRAMID PLACE MEMPHIS TN 38132 US 38132


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.