CELLEBRITY? M00516151 1615

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2016-06-06 for CELLEBRITY? M00516151 1615 manufactured by Boston Scientific - Spencer.

Event Text Entries

[46664705] (b)(4). Investigation result: analysis of the returned cellebrity cytology brush revealed multiple bends in the working length and pull wire. The wire had been broken at the end of the hypotube. The catheter was torn/ split from the luer cap, which made the handle cannula poke through a tear when handle spool was actuated. The brush bristle section was present, properly formed, and without issue. Functional analysis showed that the brush would not extend due to wire broke. The event was most likely caused by some operational or anatomical aspect of the procedure which restricted the pull wire? S ability to move freely within the catheter. Attempts to extend and retract the brush with pull wire movement restricted resulted in the broken pull wire. Therefore, the most probably root cause for this event is determined to be operational context. A review of the device history record (dhr) was performed and no deviations were found.
Patient Sequence No: 1, Text Type: N, H10


[46664706] It was reported to boston scientific corporation that a cellebrity cytology brush was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2016. According to the complainant, during the procedure, the plastic sheath kinked. The procedure was completed with a different device. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be fine. This event has been deemed a reportable event based on the investigation results; wire broke.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2016-01529
MDR Report Key5701946
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2016-06-06
Date of Report2016-05-12
Date of Event2016-03-24
Date Mfgr Received2016-05-12
Device Manufacturer Date2015-11-05
Date Added to Maude2016-06-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELLEBRITY?
Generic NameBRUSH, BIOPSY, BRONCHOSCOPE (NON-RIGID)
Product CodeBTG
Date Received2016-06-06
Returned To Mfg2016-04-13
Model NumberM00516151
Catalog Number1615
Lot Number18590449
Device Expiration Date2017-11-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.