[393581] Case had begun in cath lab #3 when the radiation shield and arm fell from above hitting the physician and then falling to the floor. The arm broke at the elbow joint of the arm support. It seems that the 50 pound lead shield places stress on this joint when it is moved at a certain angle. This has happened four times in the past. The manufacturer, mavig, came from germany and reinforced the supports. No patient or physician injury. Case was completed without further incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number570695
MDR Report Key570695
Date Received2005-01-24
Date of Report2005-01-20
Date of Event2004-12-30
Report Date2005-01-20
Date Reported to FDA2005-01-24
Date Added to Maude2005-02-08
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMAVIG SUSSPENDED LEAD SHIELD
Generic NameRADIATION SHIELD AND ARM
Product CodeKPY
Date Received2005-01-24
Model Number0T25V50
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age3 YR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key560543
ManufacturerSIEMENS MEDICAL SOLUTIONS USA, INC.
Manufacturer Address51 VALLEY STREAM PARKWAY MALVERN PA 19355 US


Patients

Patient NumberTreatmentOutcomeDate
10 2005-01-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.