SPEED-TRACT STERILE FIELD POST WITH COUPLER 11321

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-06-09 for SPEED-TRACT STERILE FIELD POST WITH COUPLER 11321 manufactured by Integra Lifesciences Corporation Oh/usa.

Event Text Entries

[46898492] On 5/18/2016 integra investigation completed. Method: failure analysis, device history evaluation results: failure analysis: the customer? S complaint noted above has been confirmed by engineering. During evaluation of the returned unit, engineering noticed that the field post would not fully clamp onto the test rail. As the field post was being clamped onto the test rail, resistance was felt in the knob of the bolt subassembly which subsequently caused the field post to seize up completely. The unit would seize up and halt any further rotational movement of the knob. There is a considerable amount of wear and tear present over the entire surface of the defective field post which is expected from regular usage over a significant period of time. Device history evaluation: device history record reviewed for this product id shows no abnormalities related to the reported failure. The device manufactured with this lot code passed all required inspection points with no associated mrr? S, variances or rework. No service history is on file for this device. Conclusion: the presence of residual contaminates within the threads of the? Bolt? And the mating? Welded post subassembly, speed-tract? Are the primary causes for the observed resistance. The presence of these foreign contaminates are obstructing the mating threads during engagement, and as a result prevents the end user from properly/fully clamping the field posts onto the bed rail as stated. It? S likely the unit was not properly lubricated during post-cleaning inspections throughout its span of usage. The ifu instructs the end user to lubricate any? Metal to metal? Threading with a surgically acceptable lubricant in an effort to reduce frictional wear and tear of the components.
Patient Sequence No: 1, Text Type: N, H10


[46898493] Customer initially reports post will not tighten to rail. On (b)(6) 2016 no further information available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2016-00090
MDR Report Key5709257
Report SourceUSER FACILITY
Date Received2016-06-09
Date of Report2016-02-09
Date Mfgr Received2016-05-18
Device Manufacturer Date2013-03-29
Date Added to Maude2016-06-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SANDRA LEE
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA LIFESCIENCES CORPORATION OH/USA
Manufacturer Street4900 CHARLEMAR DRIVE
Manufacturer CityCINCINNATI OH 45227
Manufacturer CountryUS
Manufacturer Postal Code45227
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPEED-TRACT STERILE FIELD POST WITH COUPLER
Generic NameSURGICAL RETRACTOR
Product CodeFFO
Date Received2016-06-09
Returned To Mfg2016-02-25
Catalog Number11321
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA LIFESCIENCES CORPORATION OH/USA
Manufacturer Address4900 CHARLEMAR DRIVE 4900 CHARLEMAR DRIVE CINCINNATI OH 45227 US 45227


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-09