NV LARGE BORE Y-IRRIGATION SET 06599

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2005-01-25 for NV LARGE BORE Y-IRRIGATION SET 06599 manufactured by Hospira De Costa Rica Ltd..

Event Text Entries

[347162] Patient was undergoing a transurethral resection of the prostate. The customer reported that 3l glycine bags were being consecutively spiked and replaced by the circulating nurse while glycine was delivered via gravity. Their routine is to change bags, open the flow wide open, clamp off the empty line, and spike the next bag. After changing bag number 0 or 10, the surgeon visualized a foreign fragment, and was able to remove it from the bladder without significant delay. Subsequently, the fragment was matched up with a broken section in the piercing pin tip area. There was no adverse patient effect.
Patient Sequence No: 1, Text Type: D, B5


[6316502] Patient was undergoing a transurethral resection of the prostate. The customer reported that 3l glycine bags were being consecutively spiked and replaced by the circulating nurse while glycine was delivered via gravity. Their routine is to change bags, open the flow wide open, clamp off the empty line, and spike the next bag. After changing bag number 9 or 10, the surgeon visualized a foreign fragment, and was able to remove it from the bladder without significant delay. Subsequently, the fragment was matched up with a broken section in the piercing pin tip area. There was no adverse patient effect.
Patient Sequence No: 2, Text Type: D, B5


MAUDE Entry Details

Report Number9615050-2005-00014
MDR Report Key571155
Report Source05
Date Received2005-01-25
Date of Report2005-01-04
Date of Event2005-01-04
Date Mfgr Received2005-01-04
Device Manufacturer Date2004-08-01
Date Added to Maude2005-02-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactPAUL LANDESMAN, PH.D, ASST DIR
Manufacturer Street8401 WEST 102ND STREET SUITE 300
Manufacturer CityPLEASANT PRAIRIE WI 53158
Manufacturer CountryUS
Manufacturer Postal53158
Manufacturer Phone2625776072
Manufacturer G1HOSPIRA DE COSTA RICA LTD.
Manufacturer StreetZONA FRANCA GLOBAL
Manufacturer CityLA AURORA DE HEREDIA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameNV LARGE BORE Y-IRRIGATION SET
Generic NameIRRIGATION SET
Product CodeEYN
Date Received2005-01-25
Returned To Mfg2005-01-06
Model NumberNA
Catalog Number06599
Lot Number201295H
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key561005
ManufacturerHOSPIRA DE COSTA RICA LTD.
Manufacturer AddressZONA FRANCA GLOBAL LA AURORA DE HEREDIA CS
Baseline Brand NameNV LARGE BORE Y-IRRIGATION SET
Baseline Generic NameIRRIGATION SET
Baseline Catalog No06599
Baseline IDNA
Baseline Device FamilyTRAY, IRRIGATION, STERILE
Baseline Shelf Life Contained*
Baseline Shelf Life [Months]*
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK841362
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-01-25
20 2005-01-25

© 2026 FDA.report