MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-06-13 for ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM 8065750290 manufactured by Alcon Research, Ltd. - Irvine Technology Center.
[48201657]
Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[48201658]
A customer reported that the equipment was shutting down on its own during a procedure. The procedure was completed using manual technique. There was no harm to the patient.
Patient Sequence No: 1, Text Type: D, B5
[50195360]
The system was examined and the reported event was replicated. The power supply thermistor was replaced to address the issue. The system was tested and found to meet product specifications. A review of the customer? S complaint history for the last 12 months did not show any previous complaints of this kind against the system. The system was manufactured on january 27, 2006. Based on qa assessment, the product met specifications at the time of release. The root cause of the reported event can be attributed to a nonconforming power supply thermistor. However, how or when the component became nonconforming cannot be determined the manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2028159-2016-02497 |
MDR Report Key | 5720039 |
Date Received | 2016-06-13 |
Date of Report | 2016-07-21 |
Date of Event | 2016-05-18 |
Date Mfgr Received | 2016-07-14 |
Device Manufacturer Date | 2006-01-27 |
Date Added to Maude | 2016-06-13 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. EDDIE DARTON, MD, JD |
Manufacturer Street | 6201 SOUTH FREEWAY MAIL STOP AB2-6 |
Manufacturer City | FORT WORTH TX 76134 |
Manufacturer Country | US |
Manufacturer Postal | 76134 |
Manufacturer Phone | 8175686660 |
Manufacturer G1 | ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER |
Manufacturer Street | 15800 ALTON PARKWAY |
Manufacturer City | IRVINE CA 92618 |
Manufacturer Country | US |
Manufacturer Postal Code | 92618 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM |
Generic Name | ENDOILLUMINATOR |
Product Code | MPA |
Date Received | 2016-06-13 |
Model Number | NA |
Catalog Number | 8065750290 |
Lot Number | NA |
ID Number | 1.18 - 2.18 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER |
Manufacturer Address | 15800 ALTON PARKWAY IRVINE CA 92618 US 92618 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2016-06-13 |