CONTOUR PVA MICROSPHERES M0017600450 760045S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2016-06-14 for CONTOUR PVA MICROSPHERES M0017600450 760045S manufactured by Boston Scientific - Cork.

Event Text Entries

[47311397] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[47311398] It was reported that sterility was compromised a contour pva microspheres package for an individual unit was noted to have been opened when examined. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


[55144524] Device evaluated by mfr. : the device was received with it's original pouch. A visual inspection of the returned device found that the pouch was wrapped in plastic. The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications. There were no issues identified in the device history review that potentially relate this complaint to manufacturing. The most probable root cause is considered handling damage as the event occurred without direct patient contact. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[55144525] It was reported that sterility was compromised. A contour pva microspheres package for an individual unit was noted to have been opened when examined. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2016-05416
MDR Report Key5722734
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2016-06-14
Date of Report2016-05-16
Date Mfgr Received2016-08-23
Date Added to Maude2016-06-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. LINDA LEIMER
Manufacturer StreetONE SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634941700
Manufacturer G1BOSTON SCIENTIFIC - CORK
Manufacturer StreetBUSINESS AND TECHNOLOGY PARK MODEL FARM ROAD
Manufacturer CityCORK
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCONTOUR PVA MICROSPHERES
Generic NameAGENTS, EMBOLIC, FOR TREATMENT OF UTERINE FIBROIDS
Product CodeNAJ
Date Received2016-06-14
Returned To Mfg2016-05-27
Model NumberM0017600450
Catalog Number760045S
Lot Number18200985
Device Expiration Date2018-07-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - CORK
Manufacturer AddressBUSINESS AND TECHNOLOGY PARK MODEL FARM ROAD CORK EI


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.