[56001802] Investigation conclusion: the customer's observation was not replicated in-house with retention devices. Retention devices were tested with in-house, drug-free urine. All bzo and coc results were negative at the 5 min read time and met qc specification. No false positive results were obtained. Manufacturing batch record review did not uncover any abnormalities. Root cause could not be determined based on the information provided.
Patient Sequence No: 1, Text Type: N, H10


[56001803] Customer alleging they are receiving false positive results on the panel dip devices they have always ordered. They are using an unknown lab and the confirmation results were not available for this investigation. They are having false positive results for coc and bzo. They said this is happening on random different donors. Confirmation test: results not available. The control line is present, no test line. Number of tests: 625 tests, 25 failed. No additional or patient information provided. No adverse patient sequela reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2016-00636
MDR Report Key5724799
Report SourceHEALTH PROFESSIONAL
Date Received2016-06-14
Date of Report2016-09-02
Date Mfgr Received2016-09-02
Date Added to Maude2016-06-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMULTI-DRUG ONE STEP 9 DRUG SCREEN TEST PANEL (URINE)
Generic NameDRUG OF ABUSE TEST
Product CodeDKZ
Date Received2016-06-14
Model NumberDOA-194-291-OBC12B
Lot NumberDOA6010106
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-14