STRATUS CS STAT FLUOROMETRIC ANALYZER 10444834

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2016-06-15 for STRATUS CS STAT FLUOROMETRIC ANALYZER 10444834 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[47417595] Siemens is in process of issuing an urgent field safety notice to notify all affected customers and updating existing software to resolve the issue.
Patient Sequence No: 1, Text Type: N, H10


[47417596] Siemens internally found that a testpak can be tested on the instrument regardless of the status of calibration and/or quality control if it is within the 48 hour window of lot expiry. A message is displayed next to the testpak lot number on screen alerting the customer that the testpak will expire within 48 hours. This message overwrites any other flags or messages that should be displayed when running a sample during the last 48 hours of use life. Siemens has not received any external complaint for the issue. There was no report of injury due to this issue.
Patient Sequence No: 1, Text Type: D, B5


[48007397] Siemens healthcare is currently developing software updates to address this issue. Additional data:siemens issued an urgent field safety notice (poc 16-019. A. Us) to notify all the affected customers in june 2016.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217157-2016-00065
MDR Report Key5725827
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2016-06-15
Date of Report2016-06-23
Date of Event2016-05-17
Date Mfgr Received2016-06-22
Date Added to Maude2016-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVEN ANDBERG
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693655
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street62 FLANDERS-BARTLEY RD.
Manufacturer CityFLANDERS NJ 07836
Manufacturer CountryUS
Manufacturer Postal Code07836
Single Use3
Previous Use Code3
Removal Correction Number1217157-06-17-2016-001-C
Event Type3
Type of Report3

Device Details

Brand NameSTRATUS CS STAT FLUOROMETRIC ANALYZER
Generic NameSTRATUS CS
Product CodeJHX
Date Received2016-06-15
Catalog Number10444834
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY NEWARK DE 19702 US 19702


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.