ACESSA HANDPIECE 2000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-06-15 for ACESSA HANDPIECE 2000 manufactured by Halt Medical, Inc..

Event Text Entries

[48342217] The corrective action shall focus on reinforcement of current language in the user training material/ifu with regards to advancing the device while array is deployed. Note that this was a training case; therefore the doctor was still being proctored.
Patient Sequence No: 1, Text Type: N, H10


[48342218] One of the 7 electrode array needles from the acessa handpiece (model number 2000, lot number t150101). Broke during a surgical procedure at (b)(6) with dr. (b)(6) and physician proctor, dr. (b)(6) from (b)(6). The event occurred during handpiece placement into a 10cm fibroid. It appears the handpiece was placed into the fibroid and the array deployed to 3cm. It was noted that the array was further deployed an additional 2cm. The array was retracted due to inability to optimally deploy beyond this point. The doctor withdrew the hand piece from patient to inspect. At that time, the hand piece was inspected and the ability to deploy was assessed. The array could not be deployed. A second handpiece was pulled from customer stock to proceed with the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006443171-2016-00001
MDR Report Key5728100
Date Received2016-06-15
Date of Report2016-06-06
Date of Event2016-11-05
Date Mfgr Received2016-05-11
Device Manufacturer Date2015-01-23
Date Added to Maude2016-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LOU TATEO
Manufacturer Street131 SAND CREEK RD. SUITE B
Manufacturer CityBRENTWOOD CA 94513
Manufacturer CountryUS
Manufacturer Postal94513
Manufacturer Phone9256347949
Manufacturer G1HALT MEDICAL, INC.
Manufacturer Street131 SAND CREEK RD. SUITE B
Manufacturer CityBRENTWOOD CA 94513
Manufacturer CountryUS
Manufacturer Postal Code94513
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameACESSA HANDPIECE
Generic NameACESSA HANDPIECE
Product CodeHFG
Date Received2016-06-15
Returned To Mfg2016-05-18
Model Number2000
Catalog Number2000
Lot NumberT150101
Device Expiration Date2017-01-23
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHALT MEDICAL, INC.
Manufacturer Address131 SAND CREEK RD. SUITE B BRENTWOOD CA 94513 US 94513


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.