ACESSA HANDPIECE 2000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2016-06-15 for ACESSA HANDPIECE 2000 manufactured by Halt Medical, Inc..

Event Text Entries

[47496504] The corrective action shall focus on reinforcement of current language in the user training material/ifu with regards to advancing the device while array is deployed. Note that this was a training case; therefore, the doctor was still being proctored.
Patient Sequence No: 1, Text Type: N, H10


[47496585] See mfr. Report #3006443171-2016-00001. The same doctor continued the case with a 2nd handpiece (model number 2000, lot number t150101). The 10cm fibroid was treated numerous times. During one of the ablations, it was noted in the midst of the treatment that a temperature reading was missing from needle #7. The handpiece was retracted after completion of that treatment and inspected both laparoscopically and outside of the patient for potential damage to needle #7. At that time, needle #7 was not extending along with the other 6 needles, indicating a possible breakage in the needle similar to the first handpiece. A third handpiece was pulled from customer stock (model number 2000, lot number t150101) to complete the procedure on the second fibroid. No issues were encountered with the third handpiece.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006443171-2016-00002
MDR Report Key5728101
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2016-06-15
Date of Report2016-06-06
Date of Event2016-05-11
Date Mfgr Received2016-05-11
Device Manufacturer Date2015-01-23
Date Added to Maude2016-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LOU TATEO
Manufacturer Street131 SAND CREEK RD. SUITE B
Manufacturer CityBRENTWOOD CA 94513
Manufacturer CountryUS
Manufacturer Postal94513
Manufacturer Phone9256347949
Manufacturer G1HALT MEDICAL, INC.
Manufacturer Street131 SAND CREEK RD. SUITE B
Manufacturer CityBRENTWOOD CA 94513
Manufacturer CountryUS
Manufacturer Postal Code94513
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACESSA HANDPIECE
Generic NameACESSA HANDPIECE
Product CodeHFG
Date Received2016-06-15
Returned To Mfg2016-05-18
Model Number2000
Catalog Number2000
Lot NumberT150101
Device Expiration Date2017-01-23
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHALT MEDICAL, INC.
Manufacturer Address131 SAND CREEK RD. SUITE B BRENTWOOD CA 94513 US 94513


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-15