VITROS 5600 INTEGRATED SYSTEM 6802413

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-06-23 for VITROS 5600 INTEGRATED SYSTEM 6802413 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[49157647] The investigation determined that a non-reproducible, lower than expected amon result was obtained from a quality control fluid when using a vitros 5600 integrated system. The most likely assignable cause is an instrument issue related to incubator contamination. Following maintenance actions by an ortho field engineer, vitros amon within-lab precision and within-run precision were significantly improved.
Patient Sequence No: 1, Text Type: N, H10


[49157648] A customer reported a non-reproducible, lower than expected amon result obtained on a quality control fluid processed using a vitros 5600 integrated system. Amon result = 150. 0 umol/l vs. Expected 199. 2 umol/l. Biased results of the direction and magnitude observed may lead to inappropriate physician action, should they occur undetected on patient samples. The customer made no allegation that amon patient results were affected. There was no allegation of patient harm. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319681-2016-00156
MDR Report Key5745194
Date Received2016-06-23
Date of Report2016-06-23
Date of Event2016-05-23
Date Mfgr Received2016-05-26
Device Manufacturer Date2010-06-25
Date Added to Maude2016-06-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal Code14626
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS 5600 INTEGRATED SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeJIF
Date Received2016-06-23
Catalog Number6802413
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.