[48056764] The actual sample was not returned for evaluation. Without an actual sample, a root cause has not been determined. Batch production records were reviewed and no anomalies were detected. We annually produce over (b)(4) 3-inch and (b)(4) 6-inch cotton tipped wood applicators for more than 10 years. This is the first complaint of this nature documented from this market history. Procedures are in place to inspect cotton fiber bud density. This incident is recorded in our trend analysis system and we will monitor for similar events. Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[48056765] The facility reported that the cotton tip applicator left lint on the incision site during blepharoplasty. The lint was washed from the site. No injury resulted. No further intervention was necessary.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1216735-2016-00001
MDR Report Key5745311
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2016-06-23
Date of Report2016-04-27
Date Mfgr Received2016-03-30
Device Manufacturer Date2015-08-03
Date Added to Maude2016-06-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR WILLIAM YOUNG
Manufacturer Street31 SCHOOL STREET PO BOX 149
Manufacturer CityGUILFORD ME 044430149
Manufacturer CountryUS
Manufacturer Postal044430149
Manufacturer Phone2078763311
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePURITAN COTTON TIPPED WOOD APPLICATOR
Generic NameSWAB
Product CodeKXF
Date Received2016-06-23
Model Number803-WC
Catalog Number803-WC
Lot Number18558 OR 18485
Device Expiration Date2016-08-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPURITAN MEDICAL PRODUCTS CO. LLC
Manufacturer Address31 SCHOOL STREET PO BOX 149 GUILFORD ME 044430149 US 044430149


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-06-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.