SONASTAR XS-CURVED EXTENDED HANDPIECE SONASTAR XSE CE CFSX6-H212 (MC585)

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2005-02-03 for SONASTAR XS-CURVED EXTENDED HANDPIECE SONASTAR XSE CE CFSX6-H212 (MC585) manufactured by Misonix, Inc..

Event Text Entries

[345943] Used during craniectomy for tumor, irrigation will not shut off after two minutes as described to customer. Irrigation was running continuously for approximately 1 hour but was not utilized during procedure, at that point (400 cc of fluid). Suction was on zero, however, it was suctioning. Surgeon states that it was too much. Plugged in coagulation and bovie, turned it on, unit continually ran. Unable to turn it off. They must unplug from wall. No alarms on device and ultrasonics were noted as running fine. At the time of the call the patient had not suffered any adverse effects as a result of the incident. Surgery was prolonged approximately 1 hour.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2435119-2005-00003
MDR Report Key574603
Report Source08
Date Received2005-02-03
Date of Report2005-01-31
Date of Event2004-10-07
Date Mfgr Received2004-10-11
Device Manufacturer Date2004-04-01
Date Added to Maude2005-02-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactRONALD MANNA
Manufacturer Street1938 NEW HIGHWAY
Manufacturer CityFARMINGDALE NY 11735
Manufacturer CountryUS
Manufacturer Postal11735
Manufacturer Phone6316949555
Manufacturer G1MISONIX INC.
Manufacturer Street1938 NEW HIGHWAY
Manufacturer CityFARMINGDALE NY 11735
Manufacturer CountryUS
Manufacturer Postal Code11735
Single Use3
Previous Use Code3
Removal Correction NumberC002
Event Type3
Type of Report3

Device Details

Brand NameSONASTAR XS-CURVED EXTENDED HANDPIECE
Generic NameULTRASONIC SURGICAL ASPIRATOR HANDPIECE
Product CodeFLF
Date Received2005-02-03
Model NumberSONASTAR XSE CE
Catalog NumberCFSX6-H212 (MC585)
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key564451
ManufacturerMISONIX, INC.
Manufacturer Address* FARMINGDALE NY * US
Baseline Brand NameSONASTAR XSE CURVED EXTENDED HANDPIECE
Baseline Generic NameULTRASONIC SURGICAL ASPIRATOR HANDPIECE
Baseline Model NoSONASTAR XSE CE
Baseline Catalog NoCFSX6-S212
Baseline IDMC 585
Baseline Device FamilySONASTAR ULTRASONIC SURGICAL SYSTEM
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK032690
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 2005-02-03

© 2026 FDA.report