FREE PSA 03289788190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-06-27 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[48307933] This event occurred in (b)(6). (b)(4). (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[48307934] The customer stated that they received erroneous results for one patient sample tested for total prostate-specific antigen (tpsa) and free prostate-specific antigen (fpsa). The fpsa results were higher than the tpsa results. This medwatch will refer to fpsa. Please refer to the medwatch with patient identifier (b)(6) for information related to tpsa. A prior event for the same patient was previously reported in manufacturer report number 1823260-2015-03864. The sample initially resulted as 0. 937 ng/ml when tested for tpsa on an e601 analyzer and it resulted as 6. 44 ng/ml when tested for fpsa on an e411 analyzer. The sample was repeated on (b)(6) 2016, resulting as 0. 920 ng/ml when tested for tpsa on an e601 analyzer and resulting as 6. 41 ng/ml when tested for fpsa on an e411 analyzer. The tpsa result of 0. 920 ng/ml and the fpsa result of 6. 44 ng/ml were reported outside of the laboratory. The patient was not adversely affected. The serial numbers of the used e601 and e411 analyzers were asked for, but not provided. A specific root cause could not be determined based on the provided information. The sample was requested for investigation, but there was not sufficient volume of the sample remaining for investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2016-00818
MDR Report Key5751067
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-06-27
Date of Report2016-06-27
Date of Event2016-06-08
Date Mfgr Received2016-06-09
Date Added to Maude2016-06-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2016-06-27
Model NumberNA
Catalog Number03289788190
Lot Number18843601
ID NumberNA
Device Expiration Date2017-02-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.