MONOPOLAR CABLE, ELECTROSURGICAL ACCESSORIES 600-290 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2005-02-23 for MONOPOLAR CABLE, ELECTROSURGICAL ACCESSORIES 600-290 * manufactured by Bowa Electronic Zubehor Fur Electrochiru.

Event Text Entries

[392533] The surgeon reported the cable is broken. No further information is available at this time. Additional information was received in september 2004 from the reporter. The customer reported that the monopolar cable had broken. No patient injury was reported. The cautery cord separated from the connector. There was an appearance of static on the screen before hand. When the surgeon started using the monopolar device, the cord was fine then the cord snapped in half. No patient injury or user injury was reported. There is a potential for an electrical shock to the user when disconnecting cord from the power supply. A spark or fire in an operating room has the potential to cause a serious injury such as a burn. In order to insure pt safety aorn standards recommend that during laparoscopic procedures one should: examine all electrodes for impaired insulation before use, ensure proper connection or equipment, ensure that the active electrode is not activated until it is in close proximity to the tissue, and use the lowest power setting that achieves the desired result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2430952-2005-00010
MDR Report Key575555
Date Received2005-02-23
Date of Report2004-02-23
Date Facility Aware2004-09-07
Report Date2005-02-23
Date Reported to FDA2005-02-23
Date Reported to Mfgr2005-02-23
Date Added to Maude2005-02-25
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMONOPOLAR CABLE, ELECTROSURGICAL ACCESSORIES
Generic NameENDOSCOPIC INSTRUMENT
Product CodeFFZ
Date Received2005-02-23
Model Number600-290
Catalog Number*
Lot NumberNA
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key565406
ManufacturerBOWA ELECTRONIC ZUBEHOR FUR ELECTROCHIRU
Manufacturer Address* GOMARINGEN GM


Patients

Patient NumberTreatmentOutcomeDate
10 2005-02-23

© 2024 FDA.report