TC MONITOR TCM4 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,02 report with the FDA on 2005-02-09 for TC MONITOR TCM4 * manufactured by Radiometer Medical Aps.

Event Text Entries

[393511] It has been discovered within radiometer, that a rapid switching of the instrument off and on may cause the build-in acoustical alarm to remain silent when tripped. The visual alarm remains unaffected. Radiometer has received no reports about this from the field and no harm to any pt has occurred.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2005-00001
MDR Report Key575957
Report Source01,02
Date Received2005-02-09
Date of Report2005-02-04
Date Added to Maude2005-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLEIF GUDNITZ, CONSULTANT
Manufacturer StreetAKANDEVEJ
Manufacturer CityBRONSHOJ 2700
Manufacturer CountryDA
Manufacturer Postal2700
Manufacturer Phone**
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAKANDEVEJ
Manufacturer CityBRONSHOJ 2700
Manufacturer CountryDA
Manufacturer Postal Code2700
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number3002807968-02/02/2005-00
Event Type3
Type of Report3

Device Details

Brand NameTC MONITOR
Generic NameTRANSCUTANEOUS OXYGEN AND CARBONDIOXIDE MONI
Product CodeKLK
Date Received2005-02-09
Model NumberTCM4
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key565807
ManufacturerRADIOMETER MEDICAL APS
Manufacturer Address* * *
Baseline Brand NameTC MONITOR
Baseline Generic NameTRANSCUTANEOUS OXYGEN AND CARBONDIOXIDE MONI
Baseline Model NoTCM4
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2005-02-09

© 2026 FDA.report