STERILE-DRIER 44 54344-M320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-06-29 for STERILE-DRIER 44 54344-M320 manufactured by Natus Medical Incorporated.

Event Text Entries

[49238455] Natus medical incorporated is continuing to investigate this event in order to determine the root cause of the failure. Natus is attempting to return device for evaluation, at this time it is unknown if device will return as requested by natus. A follow up report will be submitted if failed device is returned for evaluation or any additional information obtained to complete investigation. Unknown if device will be returned.
Patient Sequence No: 1, Text Type: N, H10


[49238456] Customer called and stated that sterile drier 44 device failed and caused a spark which caused considerable burns inside the dryer. Photos provided by the customer show evidence of abnormal heat in upper compartment which melted styrofoam; however photos revealed no indication of fire or burns in the vicinity.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2016-00002
MDR Report Key5760624
Date Received2016-06-29
Date of Report2016-06-29
Date Mfgr Received2016-06-01
Device Manufacturer Date2012-02-24
Date Added to Maude2016-06-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSVETLANA ROMANETS
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685170
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTERILE-DRIER 44
Generic NameSTERILE-DRIER 44
Product CodeLDS
Date Received2016-06-29
Model Number54344-M320
OperatorOTHER HEALTH CARE PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
10 2016-06-29