MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2016-07-01 for DS ANTERIOR 1.5T 9896-030-23954 manufactured by Invivo Corporation.
[48647586]
Event is under investigation and updates will be provided on a later date. (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[48647587]
A male patient was positioned head first supine and scanned for pelvis examination using the ds anterior 1. 5t coil. The patient experienced a heating sensation and 2nd degree burns resulting in blisters ( approx. 2cm x 2cm) on the inside of both inner thighs.
Patient Sequence No: 1, Text Type: D, B5
[59473587]
Based on the provided information and test performed on site, there is no indication of a malfunction of the coil used that contributed. The injuries on the inside of the patient? S inner thighs are consistent with injuries caused by so called skin-to-skin contact. Although padding was used between the legs this could not prevent the upper inner thighs from touching. It was stated that there was potential for skin-to-skin contact close to the affected area. Furthermore, 3 scans on high sar level were administered that may have contributed.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1056069-2016-00001 |
MDR Report Key | 5764841 |
Report Source | DISTRIBUTOR |
Date Received | 2016-07-01 |
Date of Report | 2016-06-28 |
Date Mfgr Received | 2016-06-01 |
Device Manufacturer Date | 2014-07-19 |
Date Added to Maude | 2016-07-01 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. KENNETH REVENNAUGH |
Manufacturer Street | 3545 SW 47TH AVE |
Manufacturer City | GAINSVILLE FL 32608 |
Manufacturer Country | US |
Manufacturer Postal | 32608 |
Manufacturer G1 | PHILIPS HEALTHCARE - INVIVO DIAGNOSTIC IMAGING |
Manufacturer Street | 3545 SW 47TH AVE |
Manufacturer City | GAINESVILLE FL 32608 |
Manufacturer Country | US |
Manufacturer Postal Code | 32608 |
Single Use | 3 |
Remedial Action | OT |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | DS ANTERIOR 1.5T |
Generic Name | DS ANTERIOR 1.5T |
Product Code | MOS |
Date Received | 2016-07-01 |
Model Number | 9896-030-23954 |
Lot Number | N/A |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | INVIVO CORPORATION |
Manufacturer Address | 3545 SW 47TH AVE GAINSVILLE FL 32608 US 32608 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2016-07-01 |