DS ANTERIOR 1.5T 9896-030-23954

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2016-07-01 for DS ANTERIOR 1.5T 9896-030-23954 manufactured by Invivo Corporation.

Event Text Entries

[48647586] Event is under investigation and updates will be provided on a later date. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[48647587] A male patient was positioned head first supine and scanned for pelvis examination using the ds anterior 1. 5t coil. The patient experienced a heating sensation and 2nd degree burns resulting in blisters ( approx. 2cm x 2cm) on the inside of both inner thighs.
Patient Sequence No: 1, Text Type: D, B5


[59473587] Based on the provided information and test performed on site, there is no indication of a malfunction of the coil used that contributed. The injuries on the inside of the patient? S inner thighs are consistent with injuries caused by so called skin-to-skin contact. Although padding was used between the legs this could not prevent the upper inner thighs from touching. It was stated that there was potential for skin-to-skin contact close to the affected area. Furthermore, 3 scans on high sar level were administered that may have contributed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056069-2016-00001
MDR Report Key5764841
Report SourceDISTRIBUTOR
Date Received2016-07-01
Date of Report2016-06-28
Date Mfgr Received2016-06-01
Device Manufacturer Date2014-07-19
Date Added to Maude2016-07-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KENNETH REVENNAUGH
Manufacturer Street3545 SW 47TH AVE
Manufacturer CityGAINSVILLE FL 32608
Manufacturer CountryUS
Manufacturer Postal32608
Manufacturer G1PHILIPS HEALTHCARE - INVIVO DIAGNOSTIC IMAGING
Manufacturer Street3545 SW 47TH AVE
Manufacturer CityGAINESVILLE FL 32608
Manufacturer CountryUS
Manufacturer Postal Code32608
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDS ANTERIOR 1.5T
Generic NameDS ANTERIOR 1.5T
Product CodeMOS
Date Received2016-07-01
Model Number9896-030-23954
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINVIVO CORPORATION
Manufacturer Address3545 SW 47TH AVE GAINSVILLE FL 32608 US 32608


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-07-01

© 2026 FDA.report