TOSHIBA MJAB-137A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-07-01 for TOSHIBA MJAB-137A manufactured by Toshiba Medical Systems Corporation.

Event Text Entries

[48664699] During a procedure a patient moved from his original position that he was placed in by the technician - arms across chest. The patient moved his arms to the side of his body which caused them to come into contact with the coil cable for remainder of the study. The patient complained of a slight burning sensation on his left inner arm after completion of the study. Patient was sent to the er. The qd torso speeder coil used was tested and coil performed correctly with no issues. It was found when checking the patient images that the patient's arm was in the field of view (fov). This indicated the improper positioning of the patient for the type of study that was being performed.
Patient Sequence No: 1, Text Type: N, H10


[48664700] Patient complained of burning sensation on arm.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020563-2016-00003
MDR Report Key5765499
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-07-01
Date of Report2016-06-07
Date of Event2016-06-03
Date Mfgr Received2016-06-07
Date Added to Maude2016-07-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PAUL BIGGINS
Manufacturer Street2441 MICHELLE DRIVE
Manufacturer CityTUSTIN CA 92780
Manufacturer CountryUS
Manufacturer Postal92780
Manufacturer Phone7147305000
Manufacturer G1TOSHIBA MEDICAL SYSTEMS CORPORATION
Manufacturer Street1385 SHIMOISHIGAMI
Manufacturer CityOTAWARA-SHI, TOCHIGI 324-8550
Manufacturer CountryJA
Manufacturer Postal Code324-8550
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOSHIBA
Generic NameMAGNETIC RESONANCE COIL
Product CodeMOS
Date Received2016-07-01
Model NumberMJAB-137A
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTOSHIBA MEDICAL SYSTEMS CORPORATION
Manufacturer Address1385 SHIMOISHIGAMI OTAWARA-SHI, TOCHIGI 324-8550 JA 324-8550


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-07-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.