RUSCH TRULED ADULT RECHARGEABLE BATTERY 0055502

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2016-07-05 for RUSCH TRULED ADULT RECHARGEABLE BATTERY 0055502 manufactured by Teleflex Medical.

Event Text Entries

[48732258] (b)(4). The device sample was not returned for evaluation at the time of this report.
Patient Sequence No: 1, Text Type: N, H10


[48732259] The customer alleges that the battery cartridge is broken. The end (metal pieces) that screws to the end of the handle is dislodged. The alleged issue was detected during pre-testing.
Patient Sequence No: 1, Text Type: D, B5


[52406496] (b)(4). The sample was returned for evaluation. A visual exam was performed and it was observed that the threaded metal ring used to secure the cartridge into the handle was missing. The cartridge could not be seated safely and securely into the truled handle. The complaint has been confirmed. The defective condition could be the result of two scenarios. The first, over torquing (tightening) of the truled cartridge threaded retaining ring. The second, dropping the truled cartridge on a hard surface and dislodging the retainer ring. While a definitive reason for either of these scenarios cannot be assigned with the information provided, it should be noted that tightening of the threaded truled cartridge can be easily achieved by using reasonable hand force, and reasonable care should be taken to ensure the battery cartridges are not dropped. A conclusion code could not be chosen as the complaint was confirmed; however, a root cause could not be established.
Patient Sequence No: 1, Text Type: N, H10


[52406497] The customer alleges that the battery cartridge is broken. The end (metal pieces) that screws to the end of the handle is dislodged. The alleged issue was detected during pre-testing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3011137372-2016-00191
MDR Report Key5768792
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2016-07-05
Date of Report2016-06-17
Date of Event2016-06-13
Date Mfgr Received2016-08-02
Date Added to Maude2016-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal Code27560
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRUSCH TRULED ADULT RECHARGEABLE BATTERY
Generic NameBATTERY, RECHARGEABLE
Product CodeFCO
Date Received2016-07-05
Returned To Mfg2016-07-14
Catalog Number0055502
Lot Number1451V3
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2016-07-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.