FLYTE LIGHTED HELMET BP 0408645000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-07-08 for FLYTE LIGHTED HELMET BP 0408645000 manufactured by Stryker Instruments-kalamazoo.

Event Text Entries

[49029828] The device was not returned for analysis.
Patient Sequence No: 1, Text Type: N, H10


[49029829] It was reported that during a surgical procedure at the user facility the device heated up. It was reported that a loose thread from the surgical hood was entangled in the fan. There was no surgical delay, no medical intervention and no adverse consequences.
Patient Sequence No: 1, Text Type: D, B5


[49503014]
Patient Sequence No: 1, Text Type: N, H10


[49503015] It was reported that during a surgical procedure at the user facility the device heated up. It was reported that a loose thread from the surgical hood was entangled in the fan. There was no surgical delay, no medical intervention and no adverse consequences.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001811755-2016-01089
MDR Report Key5778172
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-07-08
Date of Report2016-06-14
Date of Event2016-06-10
Date Mfgr Received2016-06-14
Device Manufacturer Date2013-11-04
Date Added to Maude2016-07-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CASEY METZGER
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal49001
Manufacturer Phone2693237700
Manufacturer G1STRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal Code49001
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFLYTE LIGHTED HELMET BP
Generic NameHELMET, SURGICAL
Product CodeFXZ
Date Received2016-07-08
Catalog Number0408645000
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Address4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001


Patients

Patient NumberTreatmentOutcomeDate
10 2016-07-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.