SIMPLASTIN EXCEL 252001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2005-02-24 for SIMPLASTIN EXCEL 252001 manufactured by Biomerieux, Inc..

Event Text Entries

[20990684] Simplastin excel is a thromboplastin reagent. An erroneous result was obtained on an energency room patient with simplastin excel. The result was reported to the physician. The technician thought the results were abnormally low and reconstituted a new vial and tested a second time on the patient sample. The results from the second test were immediately given to the physician. There were no adverse effects to the patient from the first reported result. The technician inspected the vial that gave the abnormal results and it ws cracked. The technician did not notice the cracked vial before the erroneius result was reported to the physician and no controls were run before the vial change-over.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2005-00004
MDR Report Key578265
Report Source07
Date Received2005-02-24
Date of Report2005-02-24
Date of Event2005-02-03
Date Mfgr Received2005-02-03
Date Added to Maude2005-03-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAMANDA KIEBY
Manufacturer Street100 RODOLPHE
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer Phone9196202053
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameSIMPLASTIN EXCEL
Generic NameTHROMBOPLASTIN REAGENT
Product CodeGGO
Date Received2005-02-24
Model NumberNA
Catalog Number252001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeNA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key568114
ManufacturerBIOMERIEUX, INC.
Manufacturer Address100 RODOLPHE ST DURHAM NC 27712 US
Baseline Brand NameSIMPLASTIN EXCEL
Baseline Generic NameTHROMBOPLASTIN REAGENT
Baseline Model NoNA
Baseline Catalog No252001
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-02-24