SURGIGUIDE 37503

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2016-07-15 for SURGIGUIDE 37503 manufactured by Dentsply Implants N.v..

Event Text Entries

[49754566] There has been a previous report received where this malfunction resulted in a serious injury. Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such. As such, this event is reportable per 21cfr part 803. Because the guide design was too fragile we redesigned the case according to our design parameters which causes the longer flanges to improve the guide strength. Based on the guide we received for complaint investigation we can conclude the clinician adjusted the flanges in order to have shorter flanges. Unfortunately we did not have any other option than making the flanges longer to avoid breakage of the guide.
Patient Sequence No: 1, Text Type: N, H10


[49754567] It was reported that a surgiguide did not properly fit in a patient's mouth. The dental implant placement surgery has been delayed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007362683-2016-00007
MDR Report Key5799084
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2016-07-15
Date of Report2016-06-15
Date of Event2016-05-03
Date Mfgr Received2016-06-15
Device Manufacturer Date2016-04-12
Date Added to Maude2016-07-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. HELEN LEWIS
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178494229
Manufacturer G1DENTSPLY IMPLANTS N.V.
Manufacturer StreetRESEARCH CAMPUS 10
Manufacturer CityHASSELT LIMBURG, B-3500
Manufacturer CountryBE
Manufacturer Postal CodeB-3500
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameSURGIGUIDE
Generic NameSYSTEM, IMAGE PROCESSING, RADIOLOGICAL
Product CodeLLZ
Date Received2016-07-15
Returned To Mfg2016-06-15
Model NumberNA
Catalog Number37503
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDENTSPLY IMPLANTS N.V.
Manufacturer AddressRESEARCH CAMPUS 10 HASSELT LIMBURG, B-3500 BE B-3500

Device Sequence Number: 2

Product CodeDZE
Date Received2016-07-15
Device Eval'ed by MfgrY
Device Sequence No2
Device Event Key0

Device Sequence Number: 3

Product CodeEBG
Date Received2016-07-15
Device Eval'ed by MfgrY
Device Sequence No3
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2016-07-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.