C1000 (TOUCH) THERMAL CYCLER CHASSIS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-07-18 for C1000 (TOUCH) THERMAL CYCLER CHASSIS manufactured by Bio-rad.

Event Text Entries

[49801286] On (b)(6) 2016, (b)(6) staff reports, heard popping coming from thermal cycler. Unit started sparking and smoking. Sparks and flames were near the power switch. Unit serial number (b)(4). Unit returned to bio-rad for repair. According to bio-rad the malfunction that caused the sparks, popping noise and flames is due to a defective power supply. Bio-rad did not issue a recall because of a less than (b)(4) failure rate, and thought that issue is contained to a power supply enclosure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5063473
MDR Report Key5802101
Date Received2016-07-18
Date of Report2016-07-12
Date of Event2016-06-08
Date Added to Maude2016-07-18
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameC1000 (TOUCH) THERMAL CYCLER CHASSIS
Generic NameTHERMAL CYCLER
Product CodeOUL
Date Received2016-07-18
Model NumberC1000
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBIO-RAD
Manufacturer Address255 LINUS PAULING DRIVE HERCULES CA 94547 US 94547


Patients

Patient NumberTreatmentOutcomeDate
10 2016-07-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.