MPS SEATING SYSTEM PS20=RB-345

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-07-19 for MPS SEATING SYSTEM PS20=RB-345 manufactured by Otto Bock Healthcare Lp.

Event Text Entries

[50729498] Per the ifu, the back adjustment was specified to be checked on a monthly basis without the user in the chair. Also, we recommend checking the product settings regularly in order to assure an optimum fit over the long term since this product is expected to be used over time. For growing children and youths in particular, fitting should be performed every six months. Also, with the child's condition it is recommended to check the chair configuration regularly to ensure proper form, fit and function.
Patient Sequence No: 1, Text Type: N, H10


[50729499] The patient became excited, jerked and flexed his body backwards applying significant force to the backrest while sitting in the chair. The backrest on the mps system broke in half and the patient fell backwards in the chair. The backrest was stabilized horizontally with the seat and held the patient above the ground in this position because the straps were in position to hold both the back and the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1721652-2016-00001
MDR Report Key5803649
Date Received2016-07-19
Date of Report2016-04-18
Date of Event2016-03-08
Date Mfgr Received2016-03-14
Device Manufacturer Date2013-12-03
Date Added to Maude2016-07-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CALEB BECK
Manufacturer Street3820 WEST GREAT LAKES DRIVE
Manufacturer CityWEST VALLEY CITY UT 84120
Manufacturer CountryUS
Manufacturer Postal84120
Manufacturer Phone8019746647
Manufacturer G1OTTO BOCK HEALTHCARE LP
Manufacturer Street3820 WEST GREAT LAKES DRIVE
Manufacturer CityWEST VALLEY CITY UT 84120
Manufacturer CountryUS
Manufacturer Postal Code84120
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMPS SEATING SYSTEM
Generic NameWHEELCHAIR COMPONENTS
Product CodeKNN
Date Received2016-07-19
Returned To Mfg2016-03-24
Model NumberPS20=RB-345
OperatorPATIENT FAMILY MEMBER OR FRIEND
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOTTO BOCK HEALTHCARE LP
Manufacturer Address3820 WEST GREAT LAKES DRIVE WEST VALLEY CITY UT 84120 US 84120


Patients

Patient NumberTreatmentOutcomeDate
10 2016-07-19