TC SENSOR 54, PCO2 945-736

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-07-20 for TC SENSOR 54, PCO2 945-736 manufactured by Radiometer Medical Aps.

Event Text Entries

[49945643] According to the complaint, an electrode burned a baby, resulting in a blister. The incident occured during a sleep study. The mother waited one week after the sleep study to report the incident to the hospital. No treatment was needed. According to the complaint, the sensor was applied for approximately 7 hours at 44 degrees. The instructions for use state that for patients aged up to (b)(6), the maximum sensor-skin interface temperature is 42 degrees celsius with a recommended maximum application time of up to 12 hours. The instructions for use state that the recommended maximum application time at 44 degrees celsius (for patients older than (b)(6) is 4 hours. Hence, the incident is caused a user error of the device.
Patient Sequence No: 1, Text Type: D, B5


[56470167] This instrument was checked out by our trained field service engineer at facility. Base unit, module and electrode were checked out and all passed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002807968-2016-00021
MDR Report Key5807962
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-07-20
Date of Report2016-10-05
Date of Event2016-06-01
Date Mfgr Received2016-09-06
Date Added to Maude2016-07-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS KRISTINA TOFT
Manufacturer StreetAAKANDEVEJ 21
Manufacturer CityBROENSHOEJ, DK-2700
Manufacturer CountryDA
Manufacturer PostalDK-2700
Manufacturer Phone538273349
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAAKANDEVEJ 21
Manufacturer CityBROENSHOEJ, DK-2700
Manufacturer CountryDA
Manufacturer Postal CodeDK-2700
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTC SENSOR 54, PCO2
Generic NameTC SENSOR 54, PCO2
Product CodeLKD
Date Received2016-07-20
Catalog Number945-736
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerRADIOMETER MEDICAL APS
Manufacturer AddressAAKANDEVEJ 21 BROENSHOEJ, BROENSHOEJ DK-2700 DA DK-2700


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-07-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.