CLINIMACS? TUBING SET LS N/A 162-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-07-20 for CLINIMACS? TUBING SET LS N/A 162-01 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[50025726]
Patient Sequence No: 1, Text Type: N, H10


[50025727] The customer reported that basically, the run started with "loading cells" but did not drain the product bag completely. Approximately half of the product went into the column. It did not rinse the pall filter and the product bag, before kicking right into the first wash. The product had severe clumping and was clogging up the fall filter. This was also creating bubbles in the drip chamber directly below the pall filter and the liquid sensor. The customer could see if the bubbles in the liquid sensor triggered the instrument while expecting that the entire product was loaded, but than it should have gone directly into rinsing the filter and bag before starting the first wash, but it did not. The customer called the miltenyi biotec inc. Hotline and the miltenyi biotec representative helped them to use another tubing set to complete the run. Therefore any risk for the patient could be ruled out.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005290010-2016-00078
MDR Report Key5809618
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-07-20
Date of Report2016-07-19
Date of Event2016-06-23
Date Mfgr Received2016-06-23
Device Manufacturer Date2015-01-08
Date Added to Maude2016-07-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS NANCY JOHANSEN
Manufacturer Street85 HAMILTON ST
Manufacturer CityCAMBRIDGE MA 02139
Manufacturer CountryUS
Manufacturer Postal02139
Manufacturer Phone6172180062
Manufacturer G1MILTENYI BIOTEC GMBH
Manufacturer StreetFRIEDRICH-EBERT-STRASSE 68
Manufacturer CityBERGISCH GLADBACH, 51429
Manufacturer CountryGM
Manufacturer Postal Code51429
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINIMACS? TUBING SET LS
Generic NameCLINIMACS? TUBING SET LS
Product CodeOVG
Date Received2016-07-20
Model NumberN/A
Catalog Number162-01
Lot NumberB2304
Device Expiration Date2017-09-02
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH