HELIOS 3000 WALL MOUNTED LIGHT HL3WL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-07-21 for HELIOS 3000 WALL MOUNTED LIGHT HL3WL manufactured by Pelton & Crane.

Event Text Entries

[51055922] Upon visual inspection by the local pelton & crane distributor, it was determined that the locking key and set screw were not properly secured on the light arm. The function of the locking key and set screw is to secure the light arm assembly together. The device was not returned to pelton & crane but rather it was serviced and placed back into use at the doctor's office and has been tested and is now functioning properly. A review of pelton & crane's helios 3000 dental light installation manual clearly documents and illustrates how to properly install the locking key and set screw during installation of the dental light. Pelton & crane reviewed the installation issue with the distributor.
Patient Sequence No: 1, Text Type: N, H10


[51055923] A medical professional was positioning a helios hl3wl for use when the light arm came apart and fell down towards the ground. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2016-00019
MDR Report Key5814055
Date Received2016-07-21
Date of Report2016-06-22
Date of Event2016-06-22
Device Manufacturer Date2012-03-13
Date Added to Maude2016-07-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1PELTON & CRANE
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHELIOS 3000 WALL MOUNTED LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2016-07-21
Model NumberHL3WL
Catalog NumberHL3WL
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2016-07-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.