ZOLL 8900-0224-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-07-20 for ZOLL 8900-0224-01 manufactured by Zoll Medical Corp..

Event Text Entries

[50463083] A patient suffered a cardiac arrest requiring defibrillation. The zoll onestep cpr complete pad had poor adherence to the patient's chest wall, even after the chest hair was shaved/removed. Remaining product from the lot number was checked and additional pads were found to be less sticky or adherent than other lots in stock. A second lot number was also found to be deficient in adherence. Both lots were removed from emergency/crash carts and replaced with effective product.
Patient Sequence No: 1, Text Type: D, B5


[50532128] Additional information received on 07/20/2016 from report or report # mw5063598: this medwatch report follows a prior report outlining issues with the adhesive properties of the zoll onestep cpr complete electrodes. Prior defective lot # reported were 2016c and 1816. Six boxes of replacement product received from (b)(4) is also defective, lacking adequate adhesive to adhere to a patient's skin. Zoll has been notified and we are expecting an overnight shipment from the company to be available on 07/21/2016.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5063598
MDR Report Key5818657
Date Received2016-07-20
Date of Report2016-07-20
Date of Event2016-07-20
Date Added to Maude2016-07-25
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameZOLL
Generic NameONESTEP CPR COMPLETE ELECTRODES
Product CodeLIX
Date Received2016-07-20
Catalog Number8900-0224-01
Lot Number2016
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No0
Device Event Key0
ManufacturerZOLL MEDICAL CORP.
Manufacturer Address269 MILL RD. CHELMSFORD MA 018244015 US 018244015

Device Sequence Number: 2

Brand NameZOLL
Generic NameONESTEP CPR COMPLETE ELECTRODES
Product CodeLIX
Date Received2016-07-20
Catalog Number8900-0224-01
Lot Number2116
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No2
Device Event Key0
ManufacturerZOLL MEDICAL CORP.
Manufacturer Address269 MILL RD. CHELMSFORD MA 018244015 US 018244015


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-07-20

© 2026 FDA.report