UTERINE CLAMP ANG 300MM MB184R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-07-27 for UTERINE CLAMP ANG 300MM MB184R manufactured by Contract Manufacturer: Tontarra Medizintechnick.

Event Text Entries

[51356814] Manufacturing site evaluation: on-going.
Patient Sequence No: 1, Text Type: N, H10


[51356815] It has been reported that the jaw of a uterine clamp broke. It has been reported that there was unknown surgical delay and unknown intervention.
Patient Sequence No: 1, Text Type: D, B5


[71636761] Investigation of the returned device indicated breakage at the area of the box hole; where the device is at its thinnest. All specifications (material and measurements) were valid at the time of production, no additional complaints have been received related to this item number. No material or manufacturing defects found. The breakage of the device is likely related to overloading or incorrect handling during use/processing.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2016-00641
MDR Report Key5828140
Date Received2016-07-27
Date of Report2017-03-30
Date Facility Aware2016-06-28
Date Mfgr Received2016-06-28
Date Added to Maude2016-07-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1CONTRACT MANUFACTURER: TONTARRA MEDIZINTECHNICK
Manufacturer Street15 DAIMLERSTR.
Manufacturer CityWURMLINGEN 78573
Manufacturer CountryGM
Manufacturer Postal Code78573
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUTERINE CLAMP ANG 300MM
Generic NameCLAMP
Product CodeHGC
Date Received2016-07-27
Model NumberMB184R
Catalog NumberMB184R
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCONTRACT MANUFACTURER: TONTARRA MEDIZINTECHNICK
Manufacturer Address15 DAIMLERSTR. WURMLINGEN 78573 GM 78573