EASY GO VAC PM65HG

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-07-28 for EASY GO VAC PM65HG manufactured by Precision Medical, Inc..

Event Text Entries

[50626371]
Patient Sequence No: 1, Text Type: N, H10


[50626372] Audit completed by biomed approximately 2 months ago - 2 other precision pm65 portable suction machines not functioning correctly, with same issue, found in 2 different units. An audit at sister hospital #1 found 2 portable suction machines not functioning correctly. No defective units found at sister hospital #2 as of 2 weeks after the initial audit. Power adapter connector was found to be broken internally preventing charging of the battery or use when plugged into an ac outlet. This indication would not be observable from the outside of the device. The connector will be replaced. During further investigation of our fleet of portable suctions: two additional devices were found with the same issue; these too will be repaired. The two suctions mentioned at sister hospital #1 were a different issue; they had the wrong voltage chargers. These will be replaced.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number5828843
MDR Report Key5828843
Date Received2016-07-28
Date of Report2016-06-28
Date of Event2016-04-17
Report Date2016-06-28
Date Reported to FDA2016-06-28
Date Reported to Mfgr2016-06-28
Date Added to Maude2016-07-28
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEASY GO VAC
Generic NamePUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)
Product CodeBTA
Date Received2016-07-28
Model NumberPM65HG
ID Number25924
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerPRECISION MEDICAL, INC.
Manufacturer Address300 HELD DRIVE NORTHAMPTON, PA 18067 US 18067


Patients

Patient NumberTreatmentOutcomeDate
10 2016-07-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.