CHROMID? CPS ELITE AGAR 416172

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-07-28 for CHROMID? CPS ELITE AGAR 416172 manufactured by Biom?rieux Sa.

Event Text Entries

[50663291] An investigation into a false negative event while using the chromid cps elite agar was performed. The customer did not return the patient samples; therefore, no testing could be performed to confirm the customer's result. Fourteen (14) quality control strains were tested on retain samples from the lot in question as well as another lot. All samples were found to be within specification. A review of the batch production record revealed no related abnormalities. Additionally, a complaint review was performed and this lot does not show an adverse quality trend. Based on the results of the investigation, the complaint could not be confirmed and the product is operating within specifications.
Patient Sequence No: 1, Text Type: N, H10


[50663292] A customer notified biomerieux that they had a discrepant result when using the chromidcps elite agar. Four (4) patient samples were run using the chromid cps elite test kit. The patient urine specimens were inoculated directly onto the plate and incubated for 24 hours. In compliance with the package insert, the customer read the plates after 24 hours and found no growth present. The customer incubated the plates for another 24 hours and upon reading the results discovered pink colonies on three (3) plates and blue-green colonies on one (1) plate. When specifically asked, the customer indicated there was no impact to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2016-00135
MDR Report Key5830129
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-07-28
Date of Report2016-06-14
Date Mfgr Received2016-06-14
Date Added to Maude2016-07-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS
Manufacturer CityCRAPONNE, FR 69290
Manufacturer CountryFR
Manufacturer Postal Code69290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? CPS ELITE AGAR
Generic NameCHROMID? CPS ELITE AGAR
Product CodeJXA
Date Received2016-07-28
Catalog Number416172
Lot Number1004811180
Device Expiration Date2016-08-16
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2016-07-28