[401669] Falselu prolonged a ptt results were obtained on 6 patient samples and were reported to the physician. Quality control samples recovered within range. The customer discovered that actin fsl and calcium chloride reagents were placed in the wrong positions on the analyzer. The actin fsl reagent was placed in the calcium chloride position and vice versa. Once reagents were placed in the correct positions, the 6 samples were re-tested. Correct qc and patient results were obtained and reported. There was no report of adverse health consequences as a result of the falsely prolonged aptt results that were initially reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2005-00003
MDR Report Key583858
Report Source05
Date Received2005-02-28
Date of Report2005-02-25
Date of Event2005-02-09
Date Mfgr Received2005-02-10
Date Added to Maude2005-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactYUK-TING LEWIS
Manufacturer StreetP.O. BOX 6101, MS 514
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026317626
Manufacturer G1DADE BEHRING MARBURG GMBH
Manufacturer Street76 EMIL VON BEHRING STRASSE
Manufacturer CityMARBURG
Manufacturer CountryGM
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Generic NamePARTIAL THROMBOPLASTIN TIME REAGENT
Product CodeGIT
Date Received2005-02-28
Model NumberNA
Catalog NumberB4219
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key573684
ManufacturerDADE BEHRING GMBH
Manufacturer Address* MARBURG GM
Baseline Brand NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Baseline Generic NamePARTIAL THROMBOPLASTIN TIME REAGENT
Baseline Model NoNA
Baseline Catalog NoB4219
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2005-02-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.