OPMI PICO ON S100 FLOOR STAND 000000-1095-102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-08-02 for OPMI PICO ON S100 FLOOR STAND 000000-1095-102 manufactured by Carl Zeiss Meditec Ag (oberkochen).

Event Text Entries

[51004062] Narrative: zeiss records indicate that there is no service contract or service history for the microscope in question. The hcp confirmed that the last service was done more than 7 years ago. A field service engineer performed an on-site inspection of the microscope and found that the pneumatic spring was leaking and did not hold position. The user manual g-30-1385-en, issue 5. 0, printed on 14. 05. 2001, advises on several pages to contact service, if a function of the microscope fails, e. G. Page 16, 38, 42. The user manual informs about a regular maintenance interval of every 24 months for the suspension system. The manufacturer's conclusion is that normal use and normal maintenance and repair following zeiss procedures could not lead to the occurred scope arm drop. The user manual advises on page 103 to perform a detailed function check before surgery and without patient present. This contains under title "setting the suspension arm weight compensation" also the proper balancing of the microscope.
Patient Sequence No: 1, Text Type: N, H10


[51004063] The healthcare professional (hcp) reported the opmi pico microscope arm was in the upper standby position when dropping suddenly about two feet. The scope head hit the patient above the left eyebrow. The hcp described the resulting abrasion as "superficial" and confirmed that no medical intervention was needed. The scope was used for routine examination of patient's ears. The examination was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2016-00012
MDR Report Key5842182
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-08-02
Date of Report2016-08-02
Date of Event2016-06-23
Date Mfgr Received2016-07-06
Device Manufacturer Date2002-07-03
Date Added to Maude2016-08-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Remedial ActionRP
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameOPMI PICO ON S100 FLOOR STAND
Generic NameMICROSCOPE, SURGICAL GENERAL&PLASTIC
Product CodeFSO
Date Received2016-08-02
Model NumberNA
Catalog Number000000-1095-102
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-08-02