OPMI PICO ON S100 WALL MOUNT 000000-1403-570

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-08-08 for OPMI PICO ON S100 WALL MOUNT 000000-1403-570 manufactured by Carl Zeiss Meditec Ag (oberkochen).

Event Text Entries

[51501027] A zeiss field service engineer (fse) performed an on-site inspection and found that the wall mount plate, on which the opmi pico s100 wall mount microscope system was mounted, came loose and caused both components, the wall mount plate and opmi pico microscope system, dropping on the floor. The fse did not find a problem with the zeiss related opmi pico s100 wall mount microscope installation onto the wall mount plate. The hcp reported that the installation of the wall mount plate was not done properly and wrong hardware (drywall screws) was used to secure the wall mount plate to the wall structure. The wall mount plate used was not a zeiss provided wall mount plate. Zeiss records confirm that the installation of the wall mount plate for preparing the wall structure was done prior to the opmi pico microscope system installation and it was confirmed to be done in accordance to the previously provided specifications defined in the zeiss planning manual. The planning manual, (b)(4), version 8. 2, (b)(6) 2014, defines the requirements for a safe mounting of the opmi pico s100 wall mount microscope system and guides the customer how to prepare the wall structure. The manual contains a complete section with "customer's preparatory responsibilities" in order ensure a safe wall structure preparation by the customer. The document advises on page 82 about the wall mount installation including a warning sign the wall mount must only be installed on the original wall flange or, if necessary, on the original wall plate from carl zeiss. "
Patient Sequence No: 1, Text Type: N, H10


[51501028] The healthcare professional (hcp) reported that the supporting wall mount plate, on which the opmi pico with s100 wall mount microscope system was mounted, came loose from the wall and the wall mount plate and the whole opmi pico microscope dropped on the floor. At the time of the incident, no one used the microsope and no patient was involved. There was no injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2016-00013
MDR Report Key5855301
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-08-08
Date of Report2016-08-06
Date of Event2016-04-01
Date Mfgr Received2016-07-20
Device Manufacturer Date2010-10-22
Date Added to Maude2016-08-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Remedial ActionIN
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameOPMI PICO ON S100 WALL MOUNT
Generic NameMICROSCOPE, SURGICAL, GENERAL & PLASTIC SURGERY
Product CodeFSO
Date Received2016-08-08
Model NumberNA
Catalog Number000000-1403-570
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-08-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.