PACEART OPTIMA POS12D16

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-08-09 for PACEART OPTIMA POS12D16 manufactured by Medtronic, Inc..

Event Text Entries

[51733808]
Patient Sequence No: 1, Text Type: N, H10


[51733809] It was reported that there was a device longevity discrepancy between the patient management database application and the programmer. The issue was escalated for investigation. The application remains in use. The status of the programmer is unknown. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2016-02044
MDR Report Key5862091
Report SourceHEALTH PROFESSIONAL
Date Received2016-08-09
Date of Report2016-06-17
Date of Event2016-06-17
Date Mfgr Received2016-06-17
Date Added to Maude2016-08-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNE SCHILLING
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635052036
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePACEART OPTIMA
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION, INDIRECT
Product CodeKRE
Date Received2016-08-09
Model NumberPOS12D16
Catalog NumberPOS12D16
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
10 2016-08-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.