TEC 7

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-08-09 for TEC 7 manufactured by Datex-ohmeda, Inc..

Event Text Entries

[51744515] Patient information could not be obtained after multiple attempts. Attempts were made as follows: 7/21/2016 - voicemail, 7/25/2016 - voicemail, 7/26/2016 - voicemail. A ge healthcare service representative performed a checkout of the system and found no issues with the anesthesia machine but confirmed an output issue with the tec 7 vaporizer. The defective vaporizer was removed from the anesthesia machine. Ge healthcare attempted to obtain the defective vaporizer for investigation but there has been no response to the requests.
Patient Sequence No: 1, Text Type: N, H10


[51744516] The hospital reported difficulty was noted while sedating a patient for a case. The hospital's agent monitor was indicating lower anesthetic agent than what was dialed in on the vaporizer. There was no reported patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2112667-2016-01524
MDR Report Key5862491
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-08-09
Date of Report2016-08-09
Date of Event2016-07-07
Date Mfgr Received2016-07-15
Device Manufacturer Date2009-04-21
Date Added to Maude2016-08-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJOSEPH SELIGA
Manufacturer Street3000 N GRANDVIEW BLVD.
Manufacturer CityWAUKESHA WI 53188
Manufacturer CountryUS
Manufacturer Postal53188
Manufacturer G1DATEX-OHMEDA, INC.
Manufacturer Street3030 OHMEDA DRIVE
Manufacturer CityMADISON WI 53718
Manufacturer CountryUS
Manufacturer Postal Code53718
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTEC 7
Generic NameVAPORIZER
Product CodeCAD
Date Received2016-08-09
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDATEX-OHMEDA, INC.
Manufacturer Address3030 OHMEDA DRIVE MADISON WI 53718 US 53718


Patients

Patient NumberTreatmentOutcomeDate
10 2016-08-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.