MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-08-10 for PALINDROME 33 CM SLOT KIT W/VT X5 8888145042P manufactured by Covidien.
[51731381]
Submit date: 8/10/2016. An investigation is currently under way; upon completion the results will be forwarded.
Patient Sequence No: 1, Text Type: N, H10
[51731382]
It was reported to covidien on (b)(6) 2016 that a customer had an issue with a dialysis catheter. The customer stated that the needle accidentally punctured through a vessel. It was used on the left internal carotid artery, and the precise needle insertion site is unknown. The procedure was cancelled, the patient was transferred by ambulance to another hospital, and is under observation. There was no product malfunction.
Patient Sequence No: 1, Text Type: D, B5
[58157098]
The actual sample involved in the reported incident was not returned for evaluation. No additional information, pictures or videos were received. Since no sample was returned for examination, it was not possible to evaluate it as part of a comprehensive failure investigation. A device history record (dhr) review revealed no discrepancies that may have contributed to a complaint of this failure mode. All quality assurance testing performed during manufacturing was acceptable. The quality assurance review of the visual, physical and dimensional evaluation results indicated that the product met specification requirements. In addition, all dhrs are reviewed for accuracy prior to product release. It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling are performed in the plant) are in place to prevent nonconforming product from leaving the manufacturing operations. This complaint will be used for tracking and trending purposes.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3009211636-2016-00347 |
MDR Report Key | 5864615 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2016-08-10 |
Date of Report | 2016-08-01 |
Date of Event | 2016-08-01 |
Date Mfgr Received | 2016-10-14 |
Date Added to Maude | 2016-08-10 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | EDWARD ALMEIDA |
Manufacturer Street | 15 HAMPSHIRE STREET |
Manufacturer City | MANSFIELD MA 02048 |
Manufacturer Country | US |
Manufacturer Postal | 02048 |
Manufacturer Phone | 5084524151 |
Manufacturer G1 | COVIDIEN |
Manufacturer Street | COVIDIEN MANUFACTURING SOLULFONS SA EDIFICIO 820 CALLE #2 ZONA FRANCA COYOL |
Manufacturer City | ALAJUELA |
Manufacturer Country | CS |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PALINDROME 33 CM SLOT KIT W/VT X5 |
Generic Name | DIALYSIS CATHETER |
Product Code | NYU |
Date Received | 2016-08-10 |
Model Number | 8888145042P |
Catalog Number | 8888145042P |
Lot Number | 1528800208 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | COVIDIEN |
Manufacturer Address | COVIDIEN MANUFACTURING SOLULFONS SA EDIFICIO 820 CALLE #2 ZONA FRANCA COYOL ALAJUELA CS |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2016-08-10 |