IVORY CALMP 12A 50057348

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-08-15 for IVORY CALMP 12A 50057348 manufactured by Heraeus Kulzer, Llc.

Event Text Entries

[53038454] Although we have not established that the device caused or contributed to the event, we're reporting it to be compliant with 21 cfr part 803 and out of an abundance of caution. Distorting the clamp during usage causing breakage of the clamp. Device breakage is addressed in the directions for use. The directions state, "do not place clamp in mouth until the rubber dam has been properly placed. Clamp could become a choking or safety hazard if dropped or broken in the mouth without proper use of the rubber dam at all times. " the directions for use warns, "caution: modification, over-extending, bending, or use exceeding one year may cause breakage. " the user used the clamp in a manner that is contraindicated by hyperextending the clamp.
Patient Sequence No: 1, Text Type: N, H10


[53038455] Broken clamp returned by (b)(6) dealer, no details.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1821514-2016-00023
MDR Report Key5875045
Date Received2016-08-15
Date of Report2016-07-27
Date Mfgr Received2016-07-27
Device Manufacturer Date2015-08-28
Date Added to Maude2016-08-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. RITA ROGERS
Manufacturer Street300 HERAEUS WAY
Manufacturer CitySOUTH BEND IN 46614
Manufacturer CountryUS
Manufacturer Postal46614
Manufacturer Phone5742995409
Manufacturer G1HERAEUS KULZER, LLC
Manufacturer Street300 HERAEUS WAY
Manufacturer CitySOUTH BEND IN 466142517
Manufacturer CountryUS
Manufacturer Postal Code466142517
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameIVORY CALMP
Generic NameCLAMP, RUBBER DAM
Product CodeEEF
Date Received2016-08-15
Returned To Mfg2016-08-10
Model Number12A
Catalog Number50057348
Lot NumberL5
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHERAEUS KULZER, LLC
Manufacturer Address300 HERAEUS WAY SOUTH BEND IN 466142517 US 466142517


Patients

Patient NumberTreatmentOutcomeDate
10 2016-08-15