STATIC AND DYNAMIC COMPLIANCE VMAX 777404-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-08-15 for STATIC AND DYNAMIC COMPLIANCE VMAX 777404-101 manufactured by Carefusion, Inc.

Event Text Entries

[52730272] The carefusion identification file is (b)(4). (b)(4). At this time, carefusion has not received the suspect device/component for evaluation. Upon completion of a final evaluation, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[52730273] The customer reported while using the static and dynamic compliance vmax device; the customer noted a burnt smell and a smoke from the unit. The customer stated the suspect device was not being used on a patient for pulmonary function testing at the time of the event, and that it was discovered during calibrations.
Patient Sequence No: 1, Text Type: D, B5


[113455512] The vyaire failure analysis lab received the suspected device and performed a failure investigation. The failure analysis technician was able to confirm the burnt capacitor (c30). The root cause is the material fatigue and is an isolated incident.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2021710-2016-04236
MDR Report Key5877350
Date Received2016-08-15
Date of Report2017-10-26
Date of Event2016-08-01
Date Mfgr Received2017-10-25
Device Manufacturer Date2006-05-01
Date Added to Maude2016-08-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149227830
Manufacturer G1CAREFUSION, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2016-08-15
Model NumberVMAX
Catalog Number777404-101
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2016-08-15

© 2024 FDA.report