CELLEBRITY? M00516151 1615

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2016-08-18 for CELLEBRITY? M00516151 1615 manufactured by Boston Scientific - Spencer.

Event Text Entries

[52503654] (b)(4). Although the suspect device has been received, the evaluation has not been completed. Therefore, the cause of the reported malfunction has not been determined. Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[52503655] It was reported to boston scientific corporation that a cellebrity cytology brush was used in the lungs during a bronchoscopy procedure performed on (b)(6) 2016. According to the complainant, during the procedure, the cytology brush was used to advance into the bronchoscope. The handle was manipulated to actuate the brush and noticed that the device was beyond the visual range of the bronchoscope, therefore, it was difficult to move the device in and out. They pulled the brush out; it was found that the plastic sheath was torn or split and the wire inside the catheter broke. The procedure was completed with another cellebrity cytology brush. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be fine.
Patient Sequence No: 1, Text Type: D, B5


[58425406] (b)(4). ? Visual evaluation of the returned device has the catheter split/torn and the pull wire was broken from the proximal end. The working length was kinked and the handle or handle cannula has no issue and was within specification. Due to anatomical/procedural factors encountered during the procedure; performance of the device was limited. Therefore, the most probable root cause classification for the reported failure is operational context. A review of the device history record (dhr) was performed and no deviations were found. A search of the complaint database revealed that no other complaints exist for this specified lot.
Patient Sequence No: 1, Text Type: N, H10


[58425407] It was reported to boston scientific corporation that a cellebrity cytology brush was used in the lungs during a bronchoscopy procedure performed on (b)(6) 2016. According to the complainant, during the procedure, the cytology brush was used to advance into the bronchoscope. The handle was manipulated to actuate the brush and noticed that the device was beyond the visual range of the bronchoscope, therefore, it was difficult to move the device in and out. They pulled the brush out; it was found that the plastic sheath was torn or split and the wire inside the catheter broke. The procedure was completed with another cellebrity cytology brush. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2016-02337
MDR Report Key5886980
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2016-08-18
Date of Report2016-07-26
Date of Event2016-07-20
Date Mfgr Received2016-09-18
Device Manufacturer Date2016-05-26
Date Added to Maude2016-08-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELLEBRITY?
Generic NameBRUSH, BIOPSY, BRONCHOSCOPE (NON-RIGID)
Product CodeBTG
Date Received2016-08-18
Returned To Mfg2016-08-05
Model NumberM00516151
Catalog Number1615
Lot Number19301691
Device Expiration Date2018-05-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2016-08-18